Abstract

Background: Multislice spiral computed tomography (MSCT) is a promising technique for noninvasive coronary angiography, although clinical application has remained limited because of frequently incomplete interpretability, caused by motion artifacts and calcifications.

Methods and results: In 59 patients (53 male, aged 58+/-12 years) with suspected obstructive coronary artery disease, ECG-gated MSCT angiography was performed with a 16-slice MSCT scanner (0.42-s rotation time, 12x0.75-mm detector collimation). Thirty-four patients were given additional beta-blockers (average heart rate: 56+/-6 min(-1)). After contrast injection, all data were acquired during an approximately 20-s breath hold. The left main (LM), left anterior descending (LAD), left circumflex (LCX), and right coronary artery (RCA), including > or =2.0-mm side branches, were independently evaluated by two blinded observers and screened for > or =50% stenoses. The consensus reading was compared with quantitative coronary angiography. MSCT was successful in 58 patients. Eighty-six of the 231 evaluated branches were significantly diseased. Without exclusion of branches, the sensitivity, specificity and positive and negative predictive value to identify > or =50% obstructed branches was 95% (82/86), 86% (125/145), 80% (82/102), and 97% (125/129), respectively. The overall accuracy for the LM, LAD, RCA, and LCX was 100%, 91%, 86%, and 81%, respectively. No obstructed LM, LAD, or RCA branches remained undetected. Classification of patients as having no, single, or multivessel disease was accurate in 78% (45/58) of patients and no patients with significant obstructions were incorrectly excluded.

Conclusions: Improvements in MSCT technology, combined with heart rate control, allow reliable noninvasive detection of obstructive coronary artery disease.

Publication types

  • Clinical Trial

MeSH terms

  • Coronary Angiography / methods*
  • Coronary Artery Disease / diagnostic imaging*
  • Coronary Artery Disease / physiopathology
  • Coronary Stenosis / diagnostic imaging
  • Female
  • Heart Rate
  • Humans
  • Kinetics
  • Male
  • Middle Aged
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / methods*